Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning
Endress+Hauser 37
Effect of the control input Only when used as control input:
triggers the allocated control function on the device when activated.
“only record”: the control input has no function, it is just recorded.
“External intermediate analysis”: as long as the control input is activated all active channels are analysed, i.e. the analysis
of min./max. averages is not made in cycles in fixed intervals / intermediate analyses, rather it is controlled via this digi-
tal input. The analysis starts when the input is activated. It is completed and saved when deactivated.
Example:
Acquiring minima, maxima, averages and quantities of a batch production.
# Warning!
Under “Signal analysis”, “Intermediate analysis” must be set to “external”.
“Lock set-up”: as long as the control input is activated, the “Set-up” key is not accessible. The set parameters can not be
displayed or changed until the control input deactivated again.
“Show text”: the specified text is shown in a message box, as long as the input is activated. If “Text, 1st part” and “Text,
2nd part” have been filled in, both parts are displayed together in the box.
Tip:
Use this, for example, as a brief procedure instruction for the operator on site.
“Display group”: the display switches to the selected group when the input is activated (signal leap from Low to High).
“Display dark”: as long as the input is activated, the display's back lighting is switched off. The device's remaining func-
tions (e.g. storage, limit value monitoring, etc.) are not affected by this.
“Synchronise time”: when the input is activated (signal leap from Low to High), the internal second counter is set to
zero. If the second indicator is between 0 and 29, the minutes remain unchanged. If the second indicator is between 30
and 59, the minutes are increased by one.
Tip:
Use this function in conjunction with a master clock. All devices used in this way run synchronously. Any signals measu-
red on different devices can be clearly compared with each other chronologically.
“Product Start/Stop”: indicates when the production of a previously selected product starts / stops. If product-specific
has been selected in “Basic settings - Operating modes - Limit values”, those limit values allocated to the selected product
are monitored as long as this digital input remains activated.
“Product selection (BCD)”: here the product is selected via a binary coded signal. For this, every digital input required
must be set to “Product selection (BCD)”. The lowest position is automatically allocated to the lowest free digital chan-
nel:
Digital input 1 = 20
Digital input 2 = 21
Digital input 3 = 2²
Digital input 4 = 2³
“Save text”: when the input is activated, a previously selected text ( see text selection (BCD) ) is saved.
“Text selection (BCD)”: here the text is selected via a binary coded signal.
“Save curve”: as long as this input is activated, the activated channels are also recorded under “Signal settings - Form
groups - Memory cycle”, i.e. the screen display and storage of the signal recording are controlled externally.
! Note!
The limit value monitoring also functions when this input has not been activated.
Use this function (input not activated), e.g. if you do not want to fill the memory with unnecessary / irrelevant measu-
rement data between batch productions or during system maintenance.
Advantage:
Better use of the internal memory and the ATA flash card, in which only really relevant information is recorded.
Set-up -> Signal settings -> Digital inputs -> Digital input x